Weight Indicator FT-10P / FT-10xx Smart Process Indicator Technical Manual Flintec GmbH Bemannsbruch 9 74909 Meckesheim GERMANY www.
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 FT-10 EN Ethernet Communication Transmission rate 10 Mbit/s, half duplex TCP
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 The meanings of the announcement LED’s on the display are: Gross Announces t
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 3.4.4 Housing FT-10 housings are panel type with stainless steel front and bac
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 4. INSTALLATION PRECAUTION: Please read this section carefully before installa
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 4.3.1 Power Supply Connection and Grounding Power supply voltage of the instr
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Junction box 6 Wire-cable FT-10 IndicatorBridges4W-Load cellShieldShield+ Exc
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Figure 4.4 – RS 485 serial interface connections Warning: Connect the shield
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 PROFIBUS Connector pin configuration (DB9F) Pin Signal Description 1 - - 2 - -
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Figure 4.10 - Cross PC connection Warning: Connect the shield to the refe
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Warning: Connect the shield to the reference ground or shield pin of the power
FT-10 Smart Process Indicator, User Manual, Rev. 1.0, January 2014 Page 1 of 92 Table Of Contents:
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 4.3.10 Digital Inputs and Outputs Connection Digital Inputs: FT-10 inputs ar
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 FT-10 AN, FT-10, FT-10 MB, FT-10 PB, FT-10 PN, FT-10 EN, FT-10 CO F
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 5. PROGRAMMING AND CALIBRATION In this section you will find the programming a
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 5.2 Fast Access to the Calibration The instrument has fast access calibration
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Changeable with calibr
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 5.4 Programming 5.4.1 Serial Port, Printer and Fieldbuses [0--] Interface
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 [01-] RS 485 Serial Interface This sub-block includes the parameters about t
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 [03-] Ethernet (Only FT-10 EN) This sub-block includes the parameters relat
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 [043 0] Print Method 0 : Printing via key. 1 : Auto print. 2 :
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 [06-] Profinet (Only FT-10 PN) This sub-block includes the parameters rela
FT-10 Smart Process Indicator, User Manual, Rev. 1.0, January 2014 Page 2 of 92
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 [12-] Filter In this block the proper filter values according to the op
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 5.4.4 Entries [14-] In this block you can enter the initial CN. [142 ]
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 [21-] Scale Build The capacity and the resolution of the scale will be defi
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 [301 ] Calibration Calibration involves emptying the scale then placing a
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 [31-] Adjustment In this sub-block you can only perform zero adjustment or
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 [313 ] eCal Calibration Warning: The scale capacity and increment shall
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 5.5.1 Analogue Output Block [4--] ( Only FT-10 AN ) The calibration of the an
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 5.5.2 Digital Output Functions [70-] In this block, the digital outputs are pr
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 [704 0] Output 5 (only FT-10, FT-10 MB, FT-10 PB, FT-10 PN, FT-10 EN and
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 5.5.4 Diagnostics [9--] The operations about checking and testing FT-10 can be
FT-10 Smart Process Indicator, User Manual, Rev. 1.0, January 2014 Page 3 of 92 RIGHTS AND LIABILITIES All rights reserved. No part of this publicati
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 6. SERIAL DATA OUTPUTS FT-10 indicator family has different kind of serial in
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 6.2 Fast Continuous Data Output Fast continuous “indicated weight” data output
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 General Rules: 1. Commands are only in CAPITAL. 2. CHK (2 ASCII char) can be
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Commands and Responses: A Read all weight data Command : [ADR][A] Resp
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 P Print :Read the stable weight Command : [ADR][P] Response : [A
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 U Read digital inputs Command : [ADR][U] Response : [ADR][U][STATUS][I
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 X Read weight value in increased resolution Command : [ADR][X] Response
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 7. ANALOGUE ( ONLY FT-10 AN ) FT-10 AN is programmable to 4 – 20 mA, 0 – 20 mA
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 [42-] Span Adjustment [420] Coarse Span Adjustment Coarse span adjustment is
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 8. MODBUS RTU ( ONLY FT-10 MB ) FT-10 MB indicator has a Modbus RTU interface
FT-10 Smart Process Indicator, User Manual, Rev. 1.0, January 2014 Page 4 of 92 1. SAFETY INSTRUCTIONS CAUTION! READ THIS MANUAL BEFORE OPERATING O
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 0 None 1 Zero 2 Tare 3 Clear 4 Print 40010 R/W 20 Not used 40030 R/W 1 Calibra
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 D1 Input-2 D2 Input-3 D3 Input-4 40201 R/W 1 Status of Outputs D0 Output-1 D1
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 4. The low byte of 40033 changes to decimal '1' at the end of the
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 9. PROFIBUS ( ONLY FT-10 PB ) In Profibus DPV1 interface, baud rate is detecte
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 9.1 GSD Configuration Profibus data consists of 2 x Input 2 words and 2 x Ou
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 FT-10 PB Output to PLC Input 2nd Dword Bit Number 2nd Dword Description D31 …
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Calibration Status (always 32 bit integer) 1st Dword descriptions when read c
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 PLC Output to FT-10 PB Input 2nd Dword Bit Number 2rd Dword descriptions D31
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Expanded Command List (always 32 bit integer) The “D23 … D16” bits in 2nd Dwo
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 10. PROFINET ( ONLY FT-10 PN ) The Profinet interface operates at 100Mbit, fu
FT-10 Smart Process Indicator, User Manual, Rev. 1.0, January 2014 Page 5 of 92 2. DECLARATION OF CONFORMITY EG-Konformitätserklärung EC-Declaration
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 10.1 Profinet Parameters There are 7 parameters for Profinet network and Prof
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 10.2 GSDML Configuration Profinet data structures of FT-10 PN includes 2 x In
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Bit Number 2nd Dword Description D31 … D24 Outputs Output bit status ( Active
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Calibration Status (always 32 bit integer) 1st Dword descriptions when read c
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 PLC Output to FT-10 PN Input 2nd Dword Bit Number 2rd Dword descriptions D31
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Expanded Command List (always 32 bit integer) The “D23 … D16” bits in 2nd Dwor
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 11. ETHERNET TCP/IP ( ONLY FT-10 EN ) Ethernet output of FT-10 EN is program
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 [036 0] Response Speed 0 : Modbus TCP Answer is sent immediately after Re
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 11.2 Modbus TCP Data Structure If the instrument is programmed for Modbus TCP
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 40031 R/W 2 Span Calibration Value / LC capacity / mV value / Dead load value
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 3. INTRODUCTION 3.1 Overview FT-10 family instruments are economic and powerfu
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 40210 R/W 2 Setpoint 3 Low ( 2 ) 40212 R/W 2 Setpoint 3 High 40214 R/W 2
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Command Examples: Performing Read and Write operations according (Modbus TCP/I
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Page 71 of 82 12. CANOPEN ( ONLY FT-10 CO ) After programming CANopen related
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Page 72 of 82 12.1 EDS Configuration CANopen data structures of FT-10 CO incl
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Page 73 of 82 12.2 CANopen Data Structure FT-10 CO Output to PLC Input Bitwi
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Page 74 of 82 (continuation of previous page) 11000 24 Set Point-5 High 11111
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Page 75 of 82 PLC Output to FT-10 CO Input RxPDO 1 (R_UL1) Bit Number RxPDO 1
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Page 76 of 82 Expanded Command List (always 32 bit integer) The “D48 … D55” bi
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Page 77 of 82 13. TROUBLE SHOOTING FT-10 weighing indicator had been designed
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Page 78 of 82 01- RS 485 206 Motion detector 2 = ± 1e 010 Data Format 5 = Mod
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 3.3 Specifications: Common Specifications A/D Converter A/D converter type 2
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Page 79 of 82 15. CALIBRATION TABLE n = Max Capacity / e recommended values
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Page 80 of 82 FREQUENTLY ASKED QUESTIONS Question : My PC could not interfa
Bemannsbruch 9 D-74909 Meckesheim Deutschland / Germany e-mail: [email protected] http:// www.flintec.com
FT-10 Smart Process Indicator, Technical Manual, Rev. 1.0, January 2014 Environment and Enclosure Operation temperature -10 °C to +40 °C; 85% RH max,